Ref. No. [UMCES] CBL 2015-008
ACT VS15-01
20
of 20.20
o
C, and the third temperature trial ranged from 30.18 to 30.54
o
C, with an average of
30.43
o
C. Time series results of the CSI, Metrohm data and reference pH data are shown in
middle panel and the time series of the calculated difference between CSI and reference pH are
shown in the bottom panel. The average difference between instrument and reference pH for the
brackish water component was 0.226 ± 0.059, with a total range of 0.121 to 0.314 (N=76).
During the freshwater trials, salinity was maintained to within 0.11 units (range = 0.19 to
0.30) with an average of 0.26 and temperature to within 0.9
o
C for each set point (Fig. 6 top).
The first temperature trial ranged from 9.96 to 10.28
o
C, with an average of 10.08
o
C, the second
temperature trial ranged from 20.00 to 20.91
o
C, with an average of 20.22
o
C, and the third
temperature trial ranged from 29.73 to 30.55
o
C, with an average of 30.42
o
C. Time series results
of the CSI, Metrohm data and reference pH data are shown in middle panel and the time series of
the calculated difference between CSI and reference pH are shown in the bottom panel. The
average difference between instrument and reference pH for the freshwater component was -
0.490 ± 0.034, with a total range of -0.569 to -0.368 (N=62).