Ref. No. [UMCES] CBL 2015-008

ACT VS15-01

20

of 20.20

o

C, and the third temperature trial ranged from 30.18 to 30.54

o

C, with an average of

30.43

o

C. Time series results of the CSI, Metrohm data and reference pH data are shown in

middle panel and the time series of the calculated difference between CSI and reference pH are

shown in the bottom panel. The average difference between instrument and reference pH for the

brackish water component was 0.226 ± 0.059, with a total range of 0.121 to 0.314 (N=76).

During the freshwater trials, salinity was maintained to within 0.11 units (range = 0.19 to

0.30) with an average of 0.26 and temperature to within 0.9

o

C for each set point (Fig. 6 top).

The first temperature trial ranged from 9.96 to 10.28

o

C, with an average of 10.08

o

C, the second

temperature trial ranged from 20.00 to 20.91

o

C, with an average of 20.22

o

C, and the third

temperature trial ranged from 29.73 to 30.55

o

C, with an average of 30.42

o

C. Time series results

of the CSI, Metrohm data and reference pH data are shown in middle panel and the time series of

the calculated difference between CSI and reference pH are shown in the bottom panel. The

average difference between instrument and reference pH for the freshwater component was -

0.490 ± 0.034, with a total range of -0.569 to -0.368 (N=62).