Background Image
Previous Page  17 / 59 Next Page
Information
Show Menu
Previous Page 17 / 59 Next Page
Page Background

Ref. No. [UMCES] CBL 2015-008

ACT VS15-01

17

comparison, but because they were calibrated directly in the test tank water against the dye

measured pH they maintained direct agreement throughout the test (Fig. 2). Linear regression of

the data has a slope of 1.006 with an r

2

of 0.999. Metrohm results are included within some of

the time series plots below to provide a more continuous record of test conditions, but no

quantification of instrument accuracy was conducted with these data.

Figure 2.

Cross plot of matched Metrohm pH results that measured continuously within the test tank at

15 minute intervals versus dye measured reference pH results. Three-hundred, fifty-one dye based pH

measurements were made over the ten week test.

Once a week multiple reference samples were collected from the test tank in a burst, i.e.

sampled as rapidly as possible (including analysis time) to examine heterogeneity within the test

tank and repeatability of the reference pH results. The average standard deviation of the

reference pH value for ten burst sampling events was 0.0025 with a range of 0.0055 to 0.0005

(Table 1).