A cross-plot of the matched observations for the first month of the deployment is given in
figure 6. Comparisons were limited to one month as a representation of a typical field deployment
and to focus on initial instrument offset versus impacts of biofouling during an extended
deployment. A linear regression of the data (r
2
= 0.99) produced a slope of 1.029 and intercept of
-0.079.
Figure 6.
CBL field response plot for EXO2 compared to reference DO samples. The plotted line
represents a 1:1 correspondence.