The global response of the AroW-USB for all nine temperature-salinity trials is shown in figure

V. In general readings are well correlated across the entire range of DO concentrations (hypoxic to

supraoxic) covering all temperature and salinity ranges tested. Highlighted deviations are likely to

reflect bubble contamination of the sensor foil. A linear regression of the data (n=257; r

2

= 0.976;

p<0.0001) produced a slope of 0.969 and intercept of 0.114. Highlighted data were omitted from the

regression because of expected bubble contamination.

Figure V

.

Concordance of JFE AroW-USB (sn 0181) oxygen measures paired with discrete Winkler

DO reference samples. Dotted line represents perfect agreement between the two measures.

Individual response slopes and intercepts of instrument versus reference DO for each of the

laboratory trials are summarized in Table 6. Response slopes increased slightly with temperature when

averaged across the three salinity trials. Means of the slopes were 0.996, 1.029, and 1.065 for 5, 15,

and 30

o

C respectively. Differences in slopes across salinity levels did not exhibit a consistent trend

and were within 0.04 across the three levels.