Ref. No. [UMCES] CBL 2016-010
ACT VS16-01
36
Figure T
. Response of JFE AroW-USB (sn 0181) to changing dissolved oxygen concentrations at 30.5
o
C and S=9.28.
Top
Panel
: Time series of instrument readings and Winkler reference samples drawn at level of sensor foil in well-mixed test
tanks. The decay in instrument signal under high DO exposure likely reflects bubble contamination of DO sensor foil.
Bottom Panel
: Corresponding offset between instrument reading and Winkler DO reference sample during the trial.
Dotted line represents zero offset from reference.