Ref. No. [UMCES] CBL 2016-015

ACT VS16-06

47

Results for the EXO2 for the two Lake Michigan trials are shown in figure 16. The

instrument exhibited a positive bias in both portions of the water column but the magnitude was

higher in the cold high DO hypolimnion. Sensor equilibration time was similar between both

trials, whether equilibrated at surface or depth. The range in measurement differences between

instrument and reference was -0.32 to -0.03 mg/L for cast 2 and -0.20 to 0.11 mg/L for cast 3.

Figure 11

. Time series of the rosette profiling and Niskin bottle sampling for the two profiling trials in

Muskegon Lake.