A cross-plot of the matched observations is given in figure 6. Comparisons were limited to
one month as a representation of a typical field deployment and to focus on initial instrument offset
versus impacts of biofouling during an extended deployment. A linear regression of the response curve
(r
2
= 0.99) produced a slope of 1.014 and intercept of -0.177.
Figure 6.
CBL field response plot a JFE AroW-USB compared to reference DO samples. The plotted line
represents a 1:1 correspondence.