ACT VS15-07

35

pH2 and reference pH over the total deployment was 0.29 ± 0.04 with a total range of 0.15 to

0.41.

Figure 16.

Time series of pH measured detected by the Xylem EXO-pH2 instrument deployed during

the 1 month MLML field trial.

Top Panel

: Continuous pH

Total

recordings from instrument (blue line) and

pH

Total

of adjacent grab samples determined by mCP dye (red circles).

Bottom Panel

: Difference in

measured pH relative to reference samples (Instrument – Dye Ref) observed during deployment. Zero

offset is represented by dashed line as pH

Total

scale is reported by both the EXO-2 sensor and the dye

reference method.