standard deviation of the difference between EXO-pH1 and reference pH over the total
deployment was 0.13 ± 0.04 with a total range of -0.01 to 0.23.
Figure 13.
Environmental conditions encountered during deployment at the MLML Small Boats Dock. Test sensor array
deployed at 1 m fixed depth, variation in local tidal heights indicate active water flow around instrument (
Top Panel
).
Variation in salinity (green) and temperature (red) at depth of instrument sensor detected by an SBE 26 and two RBR Solo
thermistors (
Middle Panel
). Temperature range determined from max-min temperatures detected by RBR and SBE
thermistors spanning instrument sensor array (
Bottom Panel
).