Ref. No. [UMCES] CBL 2015-014

ACT VS15-07

32

standard deviation of the difference between EXO-pH1 and reference pH over the total

deployment was 0.13 ± 0.04 with a total range of -0.01 to 0.23.

Figure 13.

Environmental conditions encountered during deployment at the MLML Small Boats Dock. Test sensor array

deployed at 1 m fixed depth, variation in local tidal heights indicate active water flow around instrument (

Top Panel

).

Variation in salinity (green) and temperature (red) at depth of instrument sensor detected by an SBE 26 and two RBR Solo

thermistors (

Middle Panel

). Temperature range determined from max-min temperatures detected by RBR and SBE

thermistors spanning instrument sensor array (

Bottom Panel

).