Ref. No. [UMCES] CBL 2015-013

ACT VS15-06

A cross-plot of the matched observations is given in figure 13. A linear regression was

statistically significant (p<0.001, r

2

=0.87) with a slope of 0.954. The SAI-pH under-predicted

pH for 84 of the observations and over-predicted for 17 of the observations.

Figure 13.

HIMB field response plot of a SAMI-pH(mCP) instrument compared to reference pH

Total

samples. Fine dotted line represents 1:1 correspondence, as pH

Total

scale is reported by the SAMI-pH

instrument and the dye reference method.