Ref. No. [UMCES] CBL 2015-012

ACT VS15-05

A cross-plot of the matched observations is given in figure 11. A linear regression of the

data was highly significant (p<0.001, r2 = 0.99) with a slope of 0.967 and intercept of 0.270.

Figure 11.

CBL field response plot a pumped SeaFET instrument compared to reference pH

Total

samples.

Fine dotted line represents 1:1 correspondence, as pH

Total

scale is reported by both the SeaFET instrument

and the dye reference method.