Ref. No. [UMCES] CBL 2016-011

ACT VS16-02

33

A cross-plot of all matched observations during the first month of deployment is given in

figure 3. Comparisons were limited to one month as a representation of a typical field deployment

and to focus on initial instrument offset versus impacts of biofouling during an extended

deployment. A linear regression of the instrument versus reference measurements (r

2

= 0.97;

p<0.0001) produced a slope of 0.92 and intercept of 1.03.

Figure 3

. MTU GLRI field response plot for the first month of deployment of PME miniDOT instrument

compared to reference DO samples. The plotted line represents a 1:1 correspondence. (Yellow circles

denote adjusted reference values).