Previous Page  43 / 60 Next Page
Information
Show Menu
Previous Page 43 / 60 Next Page
Page Background

Ref. No. [UMCES] CBL 2016-016

ACT VS16-07

43

A cross-plot of the matched observations for the first month of deployment is given in

figure 9. Comparisons were limited to one month as a representation of a typical field deployment

and to focus on initial instrument offset versus impacts of biofouling during an extended

deployment. A linear regression of instrument versus reference measurements for the subset data

(r

2

= 0.96, p<0.001) had a slope of 1.117 and intercept of -0.692.

Figure 9.

HIMB field response plot of Troll 9000 compared to reference DO samples determined by

Winkler Titration. The plotted line represents a1:1 correspondence.