Copyright © 2015 In-Situ Inc. This document is confidential and is the property of In-Situ Inc. Do not distribute without approval.

fully covering the measurement area, the instrument did not experience an instrument

failure but rather inability to measure due to environmental conditions.

Kaneohe Bay, HI Field Test Results: After the first month of deployment, the instrument

stopped recording data. Further analysis at the In-Situ factory showed moisture ingress that

is believed to have caused the instrument failure.

Profiling Evaluation: The high and low offsets observed with profiling evaluation are

believed to be caused by instability of the instrument temperature sensor, dissolved oxygen

sensor or a combination of both. After full stabilization, the dissolved oxygen values better

aligned with the reference instrument.

For awareness, the TROLL 9500 is a legacy product for us. The innovative scientists at In-

Situ have launched a replacement for the TROLL 9500 which is engineered for better

reliability and accuracy in harsh environmental conditions.

We are grateful to the Alliance for Coastal Technologies for inviting us to participate in this

study. The field and lab data collected and reported have provided us with an opportunity

to improve upon our product offering, training and testing procedures. We look forward

with great anticipation to participating in future studies with your organizations.

Respectfully,

David A. Bossie

Laboratory Manager and Global Safety Manager

In-Situ, Inc.