Previous Page  33 / 60 Next Page
Information
Show Menu
Previous Page 33 / 60 Next Page
Page Background

Ref. No. [UMCES] CBL 2016-016

ACT VS16-07

33

A cross-plot of the matched observations during the first month of deployment is given in

figure 3. Comparisons were limited to one month as a representation of a typical field deployment

and to focus on initial instrument offset versus impacts of biofouling during an extended

deployment. A linear regression of the data (r

2

= 0.98) produced a slope of 0.93 and intercept of

1.61, indicating an initial calibration offset.

Figure 3

. MTU GLRI field response plot for the first month of deployment of Troll 9000 DO instrument

compared to reference DO samples. The plotted line represents a 1:1 correspondence. (Yellow circles

denote adjusted reference values).