Previous Page  26 / 73 Next Page
Information
Show Menu
Previous Page 26 / 73 Next Page
Page Background

Ref. No. [UMCES] CBL 2016-010

ACT VS16-01

26

Table 5.

Summary of regression statistics for the AroUSB versus reference sample response curves for each of

the nine laboratory trials.

Test ID

R2

y Intercept

Slope

LT15S00*

0.996

-0.567

1.094

LT15S10

0.998

-0.159

0.980

LT15S35*

1.000

-0.203

1.036

LT04S00

0.999

-0.257

1.062

LT04S10

1.000

-0.173

1.038

LT04S35

1.000

-0.163

1.039

LT30S00

1.000

-0.151

1.054

LT30S10

0.995

-0.437

1.121

LT30S35

1.000

-0.105

1.065

* Bubble contamination data excluded.

Results of the 56 day long-term stability and thermal stress challenge for the AroUSB are

shown in figure K. The instrument was maintained in a well circulated tank and oxygen content

manipulated by alternately varying water temperature set point between 15 and 25

o

C several times per

week of deployment. The data completion result for the stability test was 100%. The time series of

instrument readings at 15 min intervals is plotted against discrete values for Winkler reference samples

(

top panel

) along with the time series of the difference between instrument and reference

measurements (

bottom panel

). The overall mean difference between measurements was 0.001 (s.d. =

0.326) mg/L. There was no significant trend in accuracy over time (slope = -0.0007 mg/L/d) that

would indicate any type of performance drift over the duration.

Results for the functional sensor response time assessment of the AroUSB are shown in figure

L. The top panel depicts the time series of 10s instrument reads during transfers between adjacent high

(9.6 mg/L) and low (2.0 mg/L) DO water baths, maintained commonly at 15

o

C. The bottom panel

(

lower left

) depicts results fit with a 3 parameter exponential decay function: DO

rel

= DO

relMin

+ a

e

-bt

and indicated τ calculated from fit. Data for low DO to high DO transitions (

lower right

) were treated

similarly but normalized to steady state value in subsequent high DO tank and subsequently fit with an

analogous 3 parameter exponential rise function: DO

rel

= DO

relMin

+ a(1-

e

-bt

) with indicated τ being

directly calculated from fit. The calculated τ

90

was 11.8 s during high to low transitions and 7.1 s for

low to high transitions covering a DO range of approximately 8 mg/L at a constant 15

o

C. However,

the AroUSB was incorrectly programmed for a sampling rate of 10 seconds instead of its highest

frequency of 1 s. We caution the reader that the sampling rate would have a direct impact on the

calculated response rate as noted within the manufacturer’s response.