Previous Page  32 / 59 Next Page
Information
Show Menu
Previous Page 32 / 59 Next Page
Page Background

Ref. No. [UMCES] CBL 2016-011

ACT VS16-02

32

The time series of the difference between instrument and reference DO measurements for

each matched pair (n=118 observations) is given in the bottom panel of figure 2. The average and

standard deviation of the measurement difference over the total deployment was 0.029 ± 0.072

mg/L with a total range of -0.307 to 0.205mg/L. The drift rate of instrument offset, estimated by

linear regression (r

2

=0.373; p<0.0001), was 0.001 mg/L/d. This rate would include any biofouling

effects as well as any electronic or calibration drift.

Figure 2.

Time series of DO measured dete

c

ted by PME miniDOT deployed during the 15 week Great

Lakes field trial.

Top Panel

: Continuous DO recordings from instrument (blue line) and DO of adjacent

grab samples determined by Winkler titration (red circles and yellow circles for adjusted Winkler results

due to corrections across reagent batches).

Bottom Panel

: Difference in measured DO relative to reference

samples (Instrument DO mg/L – Reference DO mg/L) observed during deployment.

Insert:

Close up of

excursion that occurred 4/8-4/9. No reference samples were collected during this time period.