Ref. No. [UMCES] CBL 2016-011

ACT VS16-02

38

A cross-plot of all matched observations for the first month of the deployment is given in

figure 6. Comparisons were limited to one month as a representation of a typical field deployment

and to focus on initial instrument offset versus impacts of biofouling during an extended

deployment. A linear regression of the instrument versus reference measurements (r

2

= 0.98;

p<0.001) produced a slope of 0.968 and intercept of 0.306.

Figure 6.

CBL field response plot for PME miniDOT compared to reference DO samples. The plotted line

represents a 1:1 correspondence.