Ref. No. [UMCES] CBL 2015-010

ACT VS15-03

A cross-plot of the matched observations is given in figure 14. The linear regression was

highly significant (p<0.001, r

2

= 0.82) and the slope of the regression was 1.55 with an intercept

of -4.90.

Figure 14.

CBL field response plot of Idronaut 305 Plus pH sensor compared to reference

pH

Total

samples. Fine dotted line represents 1:1 correspondence, coarse dotted line represents

correspondence predicted based on difference between pH

NBS

and pH

Total

in seawater (pH

NBS

=

pH

Total

+ 0.13).

36