Previous Page  44 / 44
Information
Show Menu
Previous Page 44 / 44
Page Background

mgN/L) while providing a relatively wide range (up to 10 mgN/L). In practice, this is not representative for

the performance of the instrument since two separate flow cells would be used.

4) Laboratory Testing

The ACT report states that “Instruments were set-up by the manufacturer daily prior to start of each

individual laboratory test.”. At Real Tech’s own decision, the Real Tech analyzer was not set up for each

laboratory test, but instead for the whole laboratory challenge once at the beginning. A set up performed at

the beginning of each laboratory test would have yielded better results. However, this would have been

unrealistic as real-world applications are likely to present more than one interfering factor at all times.

It is important to also note that one of the highest absolute differences for a nitrate concentration of 1 mgN/L

was observed during the concentration range test. During the range test, no interfering substance was

intended to be present in the testing tank water. Therefore, the value obtained from the range test was used

as a reference for other laboratory tests. However, all other laboratory test results for the same

concentration of 1mgN/L, with the exception of 30 PSU salinity and 10 mg/L DOC, yielded absolute

differences less than this reference value despite possible interference from the test matrix. For this reason,

we believe that the concentration range test may have been confounded by the presence of an interfering

substance in the testing tank water, despite the best efforts of the ACT team. It is possible that the

interfering substance may have affected readings of all concentrations tested during the range test.

5) Data Collection Frequency

Although data collection frequencies higher than every 15 min were not deemed an important feature for the

Challenge, we believe that in many real-world applications data collection frequencies may play a critical

role. For this reason, we would like to clarify that the accurate frequency of data readings for Real Tech’s

analyzers ranged between 2 min 34 sec and 3 min 41 sec for the Nutrient Challenge field and laboratory

tests. Moreover, data collection frequency of Real Tech’s analyzers is modifiable based on the clients’

needs and can be as high as every minute.

6) Maintenance

Real Tech’s analyzers have the capacity to alarm due to a series of potential problems that may arise and

impact the operation of the instrument. As per the Challenge’s protocol, the analyzers were not to be

serviced by the manufacturer during the Challenge as durability testing was part of the competition.

However, some of the alarms that occurred during testing could have easily been resolved by Real Tech’s

staff to ensure high accuracy data collection. Although an instrument should ideally be able to operate

without maintenance for long periods, we believe it is equally important that the instrument is intelligent

enough to inform the user of any issues that arise during deployment. This two-way communication capacity

of Real Tech’s analyzers provides an extra level of confidence for the proper operation of the instrument.

We would like to thank the ACT team again for their hard and thorough work throughout the Nutrient

Challenge. It has been an invaluable experience for Real Tech.

Sincerely,

Kerim Kollu, Ph.D.

Applications Research Scientist

Real Tech Inc.

kerim@realtechwater.com

1-905-665-6888