Ref. No. [UMCES] CBL 2016-016

ACT VS16-07

13

Thiosulfate Standardization

The Thiosulfate was standardized at room temperature as the first and last step in daily analysis.

Either triplicate assays of a fixed volume of iodate standard was run, or a range of volumes

(≥ 3) bracketing the normal sample titration range (eg. 0.500, 1.000, 1.500, 2.000 mL for well

oxygenated waters.) A clean BOD bottle and clean glassware were dedicated to this purpose.

(1)

Insert a stirring bar into a 200 mL beaker.

(2) With mixing add 1.0 mL of the H

2

SO

4

reagent followed by 1.0 mL of the alkaline iodide and

then 1.0 mL Mn

2+

reagent.

(3) Using a gravimetrically calibrated pipet add a suitable volume of the KIO

3

standard to the

stirring solution

(4) Insert the electrode and delivery tube and immediately begin titration

(5) The normality of the thiosulfate is calculated from the equivalence point volume as Vol

KIO3

/

Vol

Thio

)* N KIO

3

using replicates of single KIO

3

volume additions or from the slope of a range of

KIO

3

addition volumes.

Blank determination

Reagent blanks were determined as follows:

(1) A volume of 1-2 L of site water was brought to a boil in a clean glass reagent bottle.

(2) Boiled, degassed water was cooled and poured into 125 ml sample flasks and sparged with N

2

for no less than 30 minutes.

(3) The sample was then rapidly fixed as a normal sample, and on the auto titrator.

(4) A global reagent blank taken as the mean of samples fixed at each test site (0.078 ± 0.020, n=5)

and used to correct all reference values.