Previous Page  14 / 53 Next Page
Information
Show Menu
Previous Page 14 / 53 Next Page
Page Background

Ref. No. [UMCES] CBL 2016-012

ACT VS16-03

14

estimated as the standard deviation (s.d.) around the mean, ranged from 0.003 – 0.013 mg/L across

trials with an overall average of 0.006 mg/L. Relative precision, estimated as the coefficient of

variation (CV% = (s.d./mean)x100), ranged from 0.027 – 0.265 percent across trials with an

overall average of 0.074%.

Table 3.

Characterization of the precision of the SBS HydroCAT sensor over a range of temperatures and

salinities test conditions

.

Trial ID

Temperature

Salinity

Dissolved Oxygen Reading

mg/L mean

mg/L s.d.

CV%

n

L_T15_S00

15.44

0.00

9.163

0.005

0.054

31

L_T15_S10

15.47

8.82

9.284

0.003

0.027

31

L_T15_S35

15.39

34.20

8.145

0.004

0.053

31

L_T04_S00

5.40

0.00

12.137

0.006

0.053

31

L_T04_S10

5.30

8.98

11.830

0.007

0.057

31

L_T04_S35

5.23

34.77

10.380

0.006

0.053

31

L_T30_S00

30.22

0.00

4.856

0.013

0.265

31

L_T30_S10

30.51

9.28

6.964

0.004

0.053

31

L_T30_S35

30.61

34.43

5.915

0.003

0.047

31