Previous Page  33 / 59 Next Page
Information
Show Menu
Previous Page 33 / 59 Next Page
Page Background

Ref. No. [UMCES] CBL 2016-015

ACT VS16-06

33

A cross-plot of the matched observations during the first month of deployment is given in

figure 3. Comparisons were limited to one month as a representation of a typical field deployment

and to focus on initial instrument offset versus impacts of biofouling during an extended

deployment. A linear regression of the instrument versus reference measurements over the first

month (r

2

= 0.98; p<0.0001) produced a slope of 0.747 and intercept of 3.64 resulting from a rapid

change in response accuracy during the first few weeks.

Figure 3

. MTU GLRI field response plot for the first month of deployment of EXO2 Sonde DO instrument

compared to reference DO samples. The plotted line represents a 1:1 correspondence.