Ref. No. [UMCES] CBL 2016-015
ACT VS16-06
33
A cross-plot of the matched observations during the first month of deployment is given in
figure 3. Comparisons were limited to one month as a representation of a typical field deployment
and to focus on initial instrument offset versus impacts of biofouling during an extended
deployment. A linear regression of the instrument versus reference measurements over the first
month (r
2
= 0.98; p<0.0001) produced a slope of 0.747 and intercept of 3.64 resulting from a rapid
change in response accuracy during the first few weeks.
Figure 3
. MTU GLRI field response plot for the first month of deployment of EXO2 Sonde DO instrument
compared to reference DO samples. The plotted line represents a 1:1 correspondence.