Ref. No. [UMCES] CBL 2016-015
ACT VS16-06
32
standard deviation of the measurement difference over the total deployment was 0.37 ± 0.10 mg/L
with a total range of -0.06 to 0.56 mg/L. The drift rate of instrument offset, estimated by linear
regression (r
2
=0.85), was 0.009 mg/L/d. This rate would include any biofouling effects as well as
any electronic or calibration drift.
Figure 2.
Time series of DO measured dete
c
ted by EXO2 sonde deployed during the 15 week Great Lakes
field trial.
Top Panel
: Continuous DO recordings from instrument (blue line) and DO of adjacent grab
samples determined by Winkler titration (red circles yellow circles represent adjusted reference values, ).
Bottom Panel
: The difference in measured DO relative to reference samples (Instrument DO mg/L –
Reference DO mg/L) observed during deployment.
Insert:
Close up of excursion that occurred 4/8-4/9. No
reference samples were collected during this time period.