Ref. No. [UMCES] CBL 2016-015

ACT VS16-06

32

standard deviation of the measurement difference over the total deployment was 0.37 ± 0.10 mg/L

with a total range of -0.06 to 0.56 mg/L. The drift rate of instrument offset, estimated by linear

regression (r

2

=0.85), was 0.009 mg/L/d. This rate would include any biofouling effects as well as

any electronic or calibration drift.

Figure 2.

Time series of DO measured dete

c

ted by EXO2 sonde deployed during the 15 week Great Lakes

field trial.

Top Panel

: Continuous DO recordings from instrument (blue line) and DO of adjacent grab

samples determined by Winkler titration (red circles yellow circles represent adjusted reference values, ).

Bottom Panel

: The difference in measured DO relative to reference samples (Instrument DO mg/L –

Reference DO mg/L) observed during deployment.

Insert:

Close up of excursion that occurred 4/8-4/9. No

reference samples were collected during this time period.