Ref. No. [UMCES] CBL 2016-013
ACT VS16-04
21
for clarity. Instrument reads were adjusted to bath salinity using HOBOware’s DO Assistant
processing feature.
Figure G.
Response of the Onset HOBO logger with RDO (sn 10603120) to changing dissolved
oxygen concentrations at 30.2
o
C and S=0.00.
Top Panel
: Time series of instrument readings and
Winkler reference samples drawn at level of sensor foil in well-mixed test tanks.
Bottom Panel
:
Corresponding offset between instrument reading and Winkler DO reference sample during
exposure trial. Dotted line represents zero offset from reference.