Background Image
Previous Page  4 / 63 Next Page
Information
Show Menu
Previous Page 4 / 63 Next Page
Page Background

Ref. No. [UMCES] CBL 2015-013

ACT VS15-06

reference samples was 7.814 – 8.084 compared to a range of 7.678 – 8.220 reported by the

SAMI-pH for 4211 observations conducted continuously at 30 minute intervals. The average

and standard deviation of the measurement difference between the SAMI-pH and reference

samples over the total deployment was -0.014 ±0.016 with a total range in the differences of -

0.068 to 0.025. The useable data return for the deployment was 100%.

At Chesapeake Bay the field deployment test was conducted over 30 days with a mean

temperature and salinity of 5.9

o

C and 12.8. The measured pH range from our 107 discrete

reference samples was 8.024 – 8.403 compared to a range of 8.121 – 8.479 reported by the

SAMI-pH for 1842 observations conducted continuously at 15 minute intervals during the first

20 days of operations. Following this date, the instrument reported 175 non-determined values

and exhibited a noticeable increased offset so the remaining data was considered compromised.

The average and standard deviation of the measurement difference between the SAMI-pH and

reference samples for the period up to April 1 was 0.041 ±0.033 (n=69), with the total range of

differences between -0.086 to 0.114. The reported data return for the deployment was 94% but

again only 69% of the data was considered fully reliable.

At Lake Michigan the field deployment test was conducted over 29 days with a mean

temperature and salinity of 21.2

o

C and 0.03. Two SAMI-pH units were deployed at this test

site, one equipped with MetaCresol Purple (mCP) as the indicator dye and one with Phenol Red

(PR) as the indicator dye. The Phenol Red method was considered to be experimental for the

company at the time and was being tested for freshwater applications. The measured range in

ambient pH for the 98 reference samples was 8.013 – 8.526. The original range in pH reported

by the SAMI-pH (PR) was 6.606 – 9.305 for 2564 observations conducted continuously at 15

minute intervals. After removing fifty-one observations that were identified as outliers based on

a difference between the paired SAMI-pH units of more than two times the mean, the measured

range was reduced to 7.658 – 8.687. The average and standard deviation of the measurement

difference between the SAMI-pH (PR) and reference samples for the edited dataset was 0.021

±0.047 with a range of -0.095 to 0.177. The omitted data represents 1.7 % of the returned data

but represents a more realistic summary of the expected accuracy and uncertainty for the

technology.

The measured range of the SAMI-pH (mCP) samples was 7.813 – 8.405 based on 2665

observations conducted continuously at 15 minute intervals. The average and standard deviation

of the measurement difference between the SAMI-pH (mCP) and reference samples over the

total deployment was range was -0.162 ±0.045 and the total range of the differences was

between -0.333 to -0.068. The useable data return for this unit was 100%.

A summary plot of the SAMI-pH versus dye reference pH for all sites indicated that the

SAMI-pH responded consistently among all test sites which covered a salinity range of

freshwater to full seawater and temperatures from 3-26

o

C. As expected there was a consistent

offset in results reported within the Great Lakes test for the SAMI-pH with MetaCresol Purple

indicator dye compared with the unit operating with the Phenol Red dye indicator. The SAMI-

pH (PR) results closely match the reference samples since ACT also used Phenol Red as the

indicator dye for their analysis.

Lastly, it is worth emphasizing that the continuous 15 – 30 minute time-series provided

by the test instrument was able to resolve a significantly greater dynamic range and temporal

4