Ref. No. [UMCES] CBL 2015-010

ACT VS15-03

The 305 Plus was unresponsive to the small changes in pH during the repeated seawater

test (slope = -0.087, r

2

= 0.35) (Fig. 7). The difference between instrument and reference pH

averaged only 0.011 ±0.012 for the first 12 measurements at temperatures near 25

o

C. The

response appeared to change as temperature was cooled down to 10

o

C, and results for the next

20 matched pairs showed an offset of -0.190 ±0.108.

Figure 7.

Cross plot of Idronaut 305 Plus pH sensor measurements against the dye reference pH

measurements for 32 matched observations during the repeated seawater exposure that occurred on week

ten of the laboratory test. Temperature was varied over the course of 3 days between 26 and 10

o

C.