Ref. No. [UMCES] CBL 2016-011

ACT VS16-02

47

the profiled transitions. The range in measurement differences between instrument and reference

was -0.24 to 0.75 mg/L for cast 2 and -0.57 to 0.14 mg/L for cast 3.

Results for the PME miniDOT for the two Lake Michigan trials are shown in figure 16.

The miniDOT was well match during surface equilibration and then exhibited a strong negative

bias when rapidly transitioned to the cold high DO hypolimnion. The sensor did not fully

equilibrate after 7 minutes and ended at -0.8 mg/L against the reference. For cast 2, there was a

negative offset of -0.6 mg/L when equilibrated in the hypolimnion and a positive bias when rapidly

transitioned into the warm normoxic surface. The sensor appeared to reach equilibration after 7

minutes but with a final offset of around 0.4 mg/L against the reference. The range in

measurement differences between instrument and reference was -2.03 to 0.03 mg/L for cast 1 and -

0.72 to 1.63 mg/L for cast 2.