Ref. No. [UMCES] CBL 2016-010
ACT VS16-01
15
Table 4.
Characterization of the precision of the JFE AroW-USB (sn 0181) over a range of temperatures and
salinities.
Trial ID
Temperature
Salinity
Dissolved Oxygen Reading
mg/L mean
mg/L s.d.
CV%
n
L_T15_S00
15.44
0.00
9.035
0.004
0.040
31
L_T15_S10
15.47
8.82
9.174
0.002
0.020
31
L_T15_S35
15.39
34.20
8.035
0.002
0.030
31
L_T04_S00
5.40
0.00
12.066
0.003
0.021
31
L_T04_S10
5.30
8.98
11.783
0.005
0.040
31
L_T04_S35
5.23
34.77
10.341
0.003
0.028
31
L_T30_S00
30.22
0.00
4.798
0.012
0.247
31
L_T30_S10
30.51
9.28
6.954
0.001
0.017
31
L_T30_S35
30.61
34.43
5.877
0.001
0.018
31