Ref. No. [UMCES] CBL 2016-010

ACT VS16-01

14

Table 2.

Summary of average offset mg/L DO between paired JFE AroUSB and AroW-USB DO measures and

Winkler reference DO measures during laboratory trials.

Trial ID

AroUSB – Winkler DO

AroW-USB – Winkler DO

mean

s.d.

n

mean

s.d.

n

L_T15_S00

-0.096

0.466

42

-0.183

0.222

24

L_T15_S10

-0.277

0.245

39

-0.100

0.165

39

L_T15_S35

-0.063

0.139

32

-0.169

0.421

38

L_T04_S00

0.265

0.437

39

-0.096

0.191

39

L_T04_S10

0.141

0.237

36

-0.277

0.440

29

L_T04_S35

0.121

0.232

37

-0.145

0.114

37

L_T30_S00

0.071

0.183

40

-0.062

0.251

39

L_T30_S10

0.147

0.534

39

0.134

0.400

39

L_T30_S35

0.169

0.215

40

0.087

0.177

40

The precision of the JFE AroUSB and AroW-USB were also characterized for each of the nine

temperatures and salinity trials (Table 3 and 4, respectively). Precision trials were conducted at the

start of each new tank test when conditions were most stable. Instruments were equilibrated in test

tanks at indicated temperature and salinities for 45 min then the subsequent 31 one minute

measurements were used to estimate average tank DO (mg/L) and its variation over that interval. For

AroUSB the absolute precision, estimated as the standard deviation (s.d.) around the mean, ranged

from 0.002 – 0.014 mg/L across trials with an overall average of 0.004 mg/L. Relative precision,

estimated as the coefficient of variation (CV% = (s.d./mean)x100), ranged from 0.013 – 0.278 percent

across trials with an overall average of 0.058%.

Table 3.

Characterization of the precision of the JFE AroUSB (sn 0091) over a range of temperatures and

salinities.

Trial ID

Temperature

Salinity

Dissolved Oxygen Reading

mg/L mean

mg/L s.d.

CV%

n

L_T15_S00

15.44

0.00

9.382

0.005

0.054

31

L_T15_S10

15.47

8.82

9.516

0.002

0.023

31

L_T15_S35

15.39

34.20

8.324

0.003

0.032

31

L_T04_S00

5.40

0.00

12.485

0.002

0.013

31

L_T04_S10

5.30

8.98

12.218

0.003

0.024

31

L_T04_S35

5.23

34.77

10.747

0.004

0.035

31

L_T30_S00

30.22

0.00

4.919

0.014

0.278

31

L_T30_S10

30.51

9.28

7.063

0.002

0.033

31

L_T30_S35

30.61

34.43

5.984

0.002

0.030

31

Results for the AroW-USB were very similar (Table 4). Absolute precision ranged from 0.001

– 0.012 across trials, with an overall average of 0.004. Relative precision ranged from 0.017 – 0.247

percent across trials with an overall average of 0.051%.