Ref. No. [UMCES] CBL 2013-020
        
        
          ACT VS12-03
        
        
          Audits
        
        
          Three types of audits were performed during the verification test: a PE audit of the
        
        
          reference method measurements (GC-MS analyses), a technical systems audit (TSA) of the
        
        
          verification test performance, and a data quality audit.
        
        
          Performance Evaluation Audit
        
        
          A PE audit was conducted to assess the quality of the reference method measurements
        
        
          (GC-FID analyses) made in this verification test. The reference method PE audit was performed
        
        
          by supplying “blind” PE samples to TestAmerica, in Sacramento, CA as part of the laboratory
        
        
          tests.
        
        
          A quantitative hydrocarbon standard for TPH analysis was formulated from freshly
        
        
          opened vial of #2 Diesel Fuel Oil (5000 ppm in methanol; Spex Certiprep S-WDF-25;
        
        
          Lot#T1101213004) using a 1:10000 dilution in MilliQ water for a final concentration of 500 ppb
        
        
          directly in the sampling jars provided by Test America.  These spiked samples were shipped to
        
        
          and analyzed by Test America along with lab test samples for the same challenge compound.
        
        
          One of the spiked replicate samples was lost during transport.  For the remaining blind sample
        
        
          the Test America analysis report indicated a TPH content of 410 ppb compared to the calculated
        
        
          original concentration of 500 ppb, or an underestimate of TPH of 18% by Test America.
        
        
          Comparison of paired samples from the laboratory tests with the same lot of #2 Diesel Fuel CRM
        
        
          indicated an average underestimate of predicted TPH of 9.5%.
        
        
          Technical Systems Audit
        
        
          Two TSAs were performed during this verification.  The ACT Quality Manager
        
        
          performed a TSA on May 24-25, 2011 at Moss Landing Marine Laboratory during the initial
        
        
          laboratory tests; and the ACT Chief Scientist performed a TSA on October 31-November 2,
        
        
          2011 during the tank tests at the Bedford Institute of Oceanography.  The purpose of the TSAs
        
        
          was to assess and document the conformance of on- site testing procedures with the requirements
        
        
          of the Test Protocols and associated SOPs.  The TSAs consisted of observations of instrument
        
        
          deployments, reference sample collections and analysis, and data acquisition and handling
        
        
          procedures.  The TSAs also included an inspection of test records and documents, e.g., chain of
        
        
          custody (COC) documentation, record books, and instrument calibration logs. The audits
        
        
          confirmed that:
        
        
          
        
        
          Test instrument set-up and deployment was performed according to the Test Protocols and
        
        
          vendor instructions.
        
        
          
        
        
          Reference sample preparation procedures were performed according to the Test Protocols
        
        
          requirement.
        
        
          
        
        
          Test documentation provided a complete and traceable record of reference sample collection
        
        
          and analysis.
        
        
          
        
        
          Equipment used in the test was calibrated and monitored according to Test Protocols
        
        
          requirements and standard laboratory procedures.
        
        
          There were no adverse findings.  However, there were a number of deviations in the test
        
        
          procedures specified in the Test Protocols.  These deviations are documented in this report and
        
        
          had no negative effects on the test data quality and objectives.
        
        
          70