Background Image
Previous Page  12 / 57 Next Page
Information
Show Menu
Previous Page 12 / 57 Next Page
Page Background

Ref. No. [UMCES] CBL 2015-010

ACT VS15-03

Instrument Retrieval –

After the instruments were retrieved and cleaned of all removable fouling

according to written procedures provided by the manufacturer, a final exposure test was

conducted in the CRM buffers as defined above. For instruments that pumped samples through

an inlet, the inlet was connected to the manufacturer’s supplied storage bag to ensure that no air

was introduced prior to exposure in the buffers.

Reference Water Sampling Schedule

– The sampling frequency was structured to examine

changes in pH over daily and weekly time scales. Specifically, an intensive sampling event was

conducted once a week that consisted of 5 sample collections within a day. During four

additional days of each week, there was a minimum of twice per day sampling, scheduled in a

manner to capture as much diurnal variation as possible. The initial intensive sampling event

occurred within the first two days of the deployment after all instruments had been deployed, and

the final intensive sampling event occurred during the last two days of the deployment. The

schedule provided a higher density of comparative data at the beginning when instruments

should have been functioning at optimum performance and again after the challenge of a four or

twelve week deployment. The sampling schedule resulted in at least 60 reference samples paired

with instrument measurements. For the 12 week deployment test at HIMB, the sampling scheme

was modified to spread out a similar number of samples over the extended time period. All

sampling times were recorded on logsheets and entered into a database for final data

comparisons.

Reference Water Sample Collection

– A standard 2.2 or 4.2L Van Dorn water sampler was used

at each field test site to collect water samples for reference pH measurements. Water sample

collections were timed to correspond directly with the instrument readings or sample intake. The

water sampler was lowered to the same depth of the instrument sampling inlets, and as close as

physically possible to the inlets (distant by no more than 0.5 m horizontal distance). The water

sampler was soaked at sampling depth for 1 minute prior to sampling. If water was not flowing,

the sampler was moved to ensure that it was flushed with the ambient water. The water sampler

was triggered to match the programmed sampling times of each instrument. Three replicate pH

samples were collected in clean 10 cm glass spectrometer cuvettes from each individual field

sample. The cuvettes were gravity filled from PTFE tubing connected to the sampling bottle

spigot such that the volume of the cuvette was exchanged three times (about 20-30 secs) before a

final sample is collected. Care was taken to ensure that no bubbles were left within the cuvette

before sealing the ports with their PTFE stoppers. Cuvettes were stored at ambient or slightly

cooled conditions during transport to the laboratory (travel times at various field sites ranged

from 5 to 20 minutes) until they were placed into the temperature equilibration chamber at the

specified temperature for analysis. In addition, at the freshwater site an additional 300 ml BOD

bottle was filled to allow for a lab-based electrode pH measurement on each field reference

sample. All samples were equilibrated to 25°C and analyzed on a lab electrode calibrated daily

before use.

Twice a week (day 1 and day 5) duplicate water samples were collected to characterize

all of the CO

2

parameters. (This sample collection occurred in conjunction with our field

duplicate sampling and one set of external partner samples came from each Van Dorn. This

protocol was mostly to facilitate limitations of volume but also helped to evaluate heterogeneity

at the mooring.) Water samples were collected and preserved for pH, pCO

2

, TCO

2

and TA

measurements following standard oceanic protocols (Dickson et al., 2007). In brief, a 500 mL

glass sample bottle (supplied by Andrew Dickson, SIO) was filled from the Van Dorn sampling

12