Previous Page  13 / 59 Next Page
Information
Show Menu
Previous Page 13 / 59 Next Page
Page Background

Ref. No. [UMCES] CBL 2016-015

ACT VS16-06

13

standardization occurred within 3 - 4 s, without vortex formation. To reduce the titration time (3 -

4 min) and I

2

volatilization, an initial volume of titrant equivalent to 85–90% of the expected O

2

concentration was added at the beginning of the titration. Because the molar volume of water and

the normality of the titrant vary appreciably with temperature, care was taken to standardize the

titrant and conduct all titrations of a given batch of samples at constant temperature (± 1°C).

(1) The stopper of the BOD bottle was removed and, using a wash bottle fitted with a 200-µL

pipette tip, the I

2

present on the side and conical part of the stopper was rinsed into the BOD bottle

with 1 - 2 mL of distilled water.

(2) BOD bottles (Corning No. 5400-125) had been selected to accommodate the displacement of

the electrode without having to remove any volume of the fixed sample.

(3) The stirring bar was inserted into the bottle using plastic or stainless steel forceps.

(4) The delivery tip and the electrode were immersed, the stirrer turned on and the titration begun.

The electrode was not allowed to touch the neck of the bottle.

(5) Once the titration was complete, the equivalence point volume (

V

T) was noted

Thiosulfate Standardization

The Thiosulfate was standardized at room temperature as the first and last step in daily analysis.

Either triplicate assays of a fixed volume of iodate standard was run, or a range of volumes

(≥ 3) bracketing the normal sample titration range (eg. 0.500, 1.000, 1.500, 2.000 mL for well

oxygenated waters.) A clean BOD bottle and clean glassware were dedicated to this purpose.

(1)

Insert a stirring bar into a 200 mL beaker.

(2) With mixing add 1.0 mL of the H

2

SO

4

reagent followed by 1.0 mL of the alkaline iodide and

then 1.0 mL Mn

2+

reagent.

(3) Using a gravimetrically calibrated pipet add a suitable volume of the KIO

3

standard to the

stirring solution

(4) Insert the electrode and delivery tube and immediately begin titration

(5) The normality of the thiosulfate is calculated from the equivalence point volume as Vol

KIO3

/

Vol

Thio

)* N KIO

3

using replicates of single KIO

3

volume additions or from the slope of a range of

KIO

3

addition volumes.

Blank determination

Reagent blanks were determined as follows:

(1) A volume of 1-2 L of site water was brought to a boil in a clean glass reagent bottle.

(2) Boiled, degassed water was cooled and poured into 125 ml sample flasks and sparged with N

2

for no less than 30 minutes.

(3) The sample was then rapidly fixed as a normal sample, and on the auto titrator.

(4) A global reagent blank taken as the mean of samples fixed at each test site (0.078 ± 0.020, n=5)

and used to correct all reference values.